Mask Bacterial Filtration Efficiency (BFE) Detector GT-RA02B Product Introduction Application: Mask Bacteria Filtration Efficiency Tester,are mainly used for testing the bacterial filtration efficiency performance of masks applicable to metrological verification departments, scientific research institutes, medical mask production units and their related testing departments. It has the characteristics of fast and accurate measurement of the bacterial filtration efficiency of masks [BFE]. The Mask Bacterial Filtration Efficiency (BFE) Detector is composed of a biological aerosol generating system, aerosol chamber, aerosol delivery device, negative pressure cabinet, and 28.3LPM Anderson sampler. The entire detector is controlled by the control system. The console uses PLC+Labview to coordinate and control the work of the aerosol generation system, transmission system, negative pressure cabinet, and sampling system, and displays the working status in real time, and the entire measurement work is automatically completed. Standards:
YY 0469 Technical requirements for surgical mask BS EN 14683 Medical face masks - Requirements and test methods ASTM F2100 Standard Specification for Performance of Materials Used in Medical Face Masks ASTM F2101 Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus NBR 15052 |
Feature 1.Standard test cabinet, touch screen + PLC system is easy to operate.2.The sampler is Anderson level 6 sampler.3.The chamber is made of quartz glass with a length of 600mm and a diameter of 80mm4.The sprayer is made of quartz glass which is corrosion-resistant.5.The vacuum pump: pumping flow 57L/Min.6.The test tube holder is made of stainless steel, durable, humanized design, easy to use;7.Add fixed tooling on the base of the sampler to prevent the sampler from sliding8.The test cabinet is equipped with exhaust device to keep the negative pressure all the time.9.The door of the test chamber is equipped with large observation window and operation gloves, so as to observe during the test and operate the test.
Key Specification Model | GT-RA02B | A Route Sampling Flow | 28.3L/min, 0.01L/min | B Route Sampling Flow | 28.3L/min, 0.01L/min | Pressure of A Route Sampling Flowmeter | (-5000¢¦0)Pa | Pressure of B Route Sampling Flowmeter | (-5000¢¦0)Pa | Pressure of Spray Flowmeter | (0¢¦300)kPa | Vacuum Pump | the maximum flow rate :57LPM; Equipped with silencer, stable air flow, negative pressure: -900pa |
Peristaltic Pump Flow | (0.01¢¦3.0)mL/min | Working Temperature | 0-50¡É | Negative Pressure of Aerosol chamber | Pressure sensor range (-5000~0)Pa, negative pressure can be adjusted, normal operation negative pressure (-500~0)Pa | Negative Pressure of Main Chamber | Pressure sensor range:(-500~0)Pa, negative pressure can be adjusted by fan speed, the maximum negative pressure is about -120Pa | Data Save Ability | Data can be downloaded via U disk with a storage capacity of up to 8G | High Efficiency Particulate Air Filter Properties | ¡Ã99.99%@0.3¥ìm, | Median diameter of aerosol generator mass | Average diameter: (3.0¡¾0.3)¥ìm; Geometric Standard Deviation¡Â1.5 | Double Routes 6-lever Andersen Sampling apparatus | Level¥°£¾7¥ìm; Level ¥±: (4.7¢¦7)¥ìm; Level ¥²: (3.3¢¦4.7)¥ìm; Level ¥³: (2.1¢¦3.3)¥ìm; Level ¥´: (1.1¢¦2.1)¥ìm; Level ¥µ: (0.6¢¦1.1)¥ìm | Size of Aerosol Chamber | 600¡¿80¡¿5mm (Length¡¿Diameter¡¿Thickness) | Total number of positive quality control sampler particles | (2200¡¾500)cfu | Negative Chamber Flow Speed | ¡Ã5m3/min | Internal dimensions of operation Chamber | 740¡¿940¡¿440mm (L¡¿W¡¿H) | Size of Main Machine | 1200¡¿650¡¿2100mm (L¡¿W¡¿H) | Power Supply | AC220V¡¾10%, 50Hz | Noise of the Tester | £¼65dB(A) | Weight | ≈200kg | Power Consumption | £¼1500W |
Accessories: Standard Accessories | 1set | Air compressor | 12pcs | Petri dish | 1set | Test tube | 3m | Air pipe | 1pc | Measuring cup | 1pc | U disk |
|
|
|
|
|
|